Laminar Flow Pass Box

Laminar Flow Pass Box
Laminar Flow Pass Box is mainly used as materials in the biological transmission of clean areas, the main application areas: pharmaceuticals, scientific research units, CDC, major hospitals, university research, and application of bio-clean areas clean.

The laminar flow within the transfer window cleanliness requirements: B grade;
Inside and outside the double-shell, the internal four weeks arc process, to ensure seamless connection;
The use of laminar flow design, the air flow used to send on the next time mode, the bottom of the 304 stainless steel cold-rolled plate punching design, and set the ribs;
Filter: Primary Filter using G4, high efficiency filter used H14;
Wind speed: filter out through the high wind velocity control in the 0.38-0.57m / s (the wind in the high-flow plate are 150mm at the bottom of the test);
The pressure function: Displays the filter differential pressure (range and efficient 0-500Pa / in effect 0-250Pa), accuracy ± 5Pa;
Control: fan start / stop button to configure built-in electronic door interlock; set UV lights, design a separate switch, when The two door, the UV lamp should be turned out; set lighting, design a separate switch;
High efficiency filter can separate from the upper cabinet for removal and installation, easy maintenance and replacement of filters;
The lower part of the transfer window set in the manhole, for fans of the overhaul;
The noise: noise during normal operation of the transfer window <65db;
Both the wind and efficient flow plate: 304 stainless steel stencil.